Table. 3. Uncertainty calculated from each uncertainty factors
Uncertainty factors Combined standard uncertainty Relative standard uncertainty Degree of freedom
Dilution process for reference materials Dilution 1X 1/1 mL 0.00068 0.00068 167
Dilution 10X 0.1/0.1 mL 0.00014 0.00140 0.0016 23
0.9/1 mL 0.00071 0.00079
Dilution 100X 0.1/0.1 mL 0.00014 0.00140 0.0016 23
0.9/1 mL 0.00071 0.00079
Dilution 500X 0.4/1 mL 0.00051 0.00128 0.0017 156
0.8/1 mL 0.00064 0.00080
0.8/1 mL 0.00064 0.00080
Dilution 1000X 1/1 mL 0.00068 0.00068 0.0010 333
1/1 mL 0.00068 0.00068
Dilution 2000X 1/1 mL 0.00068 0.00068 0.0010 333
1/1 mL 0.00068 0.00068
Combined - 0.0032 165
Dilution process for samples Dilution 10X 0.1/0.1 mL 0.00014 0.00140 0.0016 23
0.9/1 mL 0.00071 0.00079
Dilution 100X 0.1/0.1 mL 0.00014 0.00140 0.0016 23
0.9/1 mL 0.00071 0.00079
Dilution 1000X 0.2/0.2 mL 0.00020 0.00099 0.0015 104
0.9/1 mL 0.00071 0.00079
0.9/1 mL 0.00071 0.00079
Dilution 2000X 1/1 mL 0.00068 0.00068 0.0010 333
1/1 mL 0.00068 0.00068
Dilution 3000X 1/1 mL 0.00068 0.00068 0.0015 39
0.5/1 mL 0.00068 0.00135
Dilution 4500X 0.6/1 mL 0.00056 0.00093 0.0019 74
0.3/1 mL 0.00051 0.00169
Combined - 0.0038 223
Calibration curve measurement using reference materials 1.091 0.0088 4
Sample measurement 2.187 0.0062 35
Yakhak Hoeji 2024;68:294-301 https://doi.org/10.17480/psk.2024.68.4.294
© 2024 Yakhak Hoeji